| Assay method: |
Absorbance, concentration, factor, 1 point end point, 2 point end (Sample blank), fixed time, kinetic, and dichromatic, immunoturbidimetric. Coagulation (optional). |
| Wave Length: |
7 filters, 340, 405, 450, 505, 546, 578, 630 as standard configuration. (5 years warranty for each filter) |
| Light source: |
6V 10W Long-life Tungsten halogen lamp. Service life 2000 hours.
Light has auto shutdown function that will extend usage period.
Auto-Sleep function. |
| Display: |
7.0 inch TFT Color touch screen,260000 Pixel,800×480 resolution; real time to show response curve |
| Built-in Thermal Printer: |
Built-in Thermal Printer, 57mm printer paper
Paper alarm, transparent cover; easy installation |
| Incubation temperature: |
37℃ and room temperature are standby; temperature accuracy±0.1℃,
by means of Peltier elements |
| Flow cell capacity and optical path: |
32ul, 10mm |
| Resolution: |
0.0001 Abs |
| Photometric Range: |
0 – 3.5 Abs |
| Absorbance range: |
0.0000-2.5000(10mm flowcell)
0.0000-3.5000(6mm flowcell) |
| Spectral Band: |
< 8 nm
|
|
Vacuum system programmable:
|
200 – 5000 ul at least |
| Carry over: |
<1%; |
| Stability: |
<0.005Abs/hour |
| Absorbance Precision: |
0.0001Abs; |
| Absorbance drift: |
<0.005Abs/hour |
| Absorbance repeatability: |
CV≤1% |
| QC analysis: |
L-J QC chart , Westguard Multi-regulation analysis. |
| Software: |
Window CE 6.0 operation system,high-capacity storage; edit 200 assay program,200000 results
Unlimited storage by connecting to computer.
Single chip machine, does not require Windows XP or Windows 7. |
| Operation Manual: |
Available in 4 different languages |
| CPU: |
ARM Cortex-A8, 720MHz |
| Memory: |
DDR2 SDRAM: 1G bit
Flash: 256 MB
Over 200 programs, 200,000 results can be stored |
| USB Port: |
USB Mouse; USB keypad; USB Printer; 4 USB ports for Keyboard, Mouse, Printer, U Disk, Bar Code Reader. 1 USB Slave for PC connections (supports PC software control function).
1 Ethernet port.
Supports all HP series inkjet printers; Computer control software,LIS system available. External scanner is available. Support remote diagnostics |
| Operational Environment: |
Temperature: 10℃~37℃. Humidity: ≤85%. |
| Power supply: |
Wide power supply: 100-240V 50Hz or 60Hz. |
| Dimensions: |
400mm (L) x 300mm (W) 160mm (H) |
| Weight: |
17 lbs.. |